Error ID: 22015, 29176
The pressure has dropped below the lower limit.
Parameter: None
Probable cause | Suggested actions | |
---|---|---|
1 | Leak | Check for leaks. |
2 | Bottle empty | Check bottle filling. Fill solvent bottle. |
3 | Wrong solvent (viscosity) | Check solvent. |
4 | Inappropriate setting | Check flow rate and lower pressure limit. |
5 | Column degradation | Replace column. |