Error ID: 22032, 29220
Pressure too far above the limit.
Probable cause | Suggested actions | |
---|---|---|
1 | Leak in degasser chamber or degasser tubing. | Please contact your Agilent service representative. |
2 | Defect vacuum pump. | Please contact your Agilent service representative. |
3 | Degasser chamber empty or connected to air. | Block unused degasser channels. |
base-id: 4202455307