When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| Spanner-double open ended 14 mm |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4220-60028 | Oulet valve (primary pump head) |
1 | ![]() | G4220-20020 | Internal gold seal for Outlet Valve |
Prerequisites
|
Remove the cap from the outlet valve.
Unscrew the outlet valve with a 14 mm wrench (1) and remove it (2).
Do not fix the screws at this stage!
Insert the outlet valve into the pump head (1). Using a torque wrench, set approx. 10 Nm and close the outlet valve (2).
Place the cap on the Outlet Valve.
Install the pump head assembly, reconnect all hydraulic connections, and power up the pump.
Open the shutoff valves and purge the system to remove air.