If a secondary leak is found, the following reasons are possible:
Probable cause | Suggested actions | |
---|---|---|
1 | Leak between the OBV and automatic purge valve |
Check for visible leaks on fittings and connectors: Connector not fixed/tight enough Connector damaged Leaky filter frit assembly
Remove the seal wash tubes from the support ring and check for leak into the seal wash path: Main seal leaking/damaged Piston damaged
Remove waste lines from the automatic purge valve: Damage to rotor seal and/or stator head |
2 | Outlet valve not properly assembled |
Check if outlet valve is properly assembled: Re-tighten the outlet valve Check the position of the gold seal |