Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5023-0271 | Frit 0.3 µm for inline filter, 5/pk |
1 | ![]() | 5230-0022 | Frit 0.3 µm for Inline Filter Biocompatible |
CAUTION
Stuck Capillary in Multi Purpose Valve
Shortcutting the inline filter by directly connecting its right capillary to valve port 5 can damage the Multi Purpose Valve.
The size/position of this capillary in its fitting is incompatible, so it may get stuck irreversibly to the valve.
NOTE
The inline filter can be cleaned using the back-flush function in the user interface of your Agilent instrument control software.
Only use the back-flush option, if an inline filter is installed.
Use two 5/16 wrenches for opening the inline filter.
Replace the filter frit and reassemble the inline filter.
Put the inline filter to the clamp and install its capillaries.