When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 9301-0411 | Syringe, Plastic |
1 | ![]() | 9301-1337 | Syringe adapter |
1 | ![]() | G7104-44000 | Transport protection foam (G7104A/C) |
1 | ![]() | G7131-98000 | Transport protection foam (G7131A/C) |
CAUTION
Mechanical damage
Flush all solvent channels with isopropanol.
Remove solvent inlet tubes from solvent reservoirs and tubing clips at other modules.
Remove tubings between the seal wash function and solvent bottle/waste.
Turn off the pump.
Remove cable and capillary connections to other modules.
Remove the waste tube.
Remove the module from the stack.
Disconnect the degasser outlet tubings at the MCGV one after another. Use a syringe for removing liquid from the degasser and the solvent tubings.
Reconnect the degasser outlet tubings to the MCGV. Remove the degasser inlet tubings.
You may keep internal tubing and capillary connections.
Carefully insert the protective foam to the front part of the instrument. Do not damage any tubing or capillary connections.
Close the doors.
For transport or shipment, put the module and accessory kit to the original shipment box.