The special settings are required for specific actions (normally in a service case).
NOTE
The tables include both settings for modules – with on-board LAN and without on-board LAN. They are identified as LAN and no LAN.
Firmware update procedures may require this mode in case of firmware loading errors (main firmware part).
If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident mode. It is not operable as a module. It only uses basic functions of the operating system for example, for communication. In this mode the main firmware can be loaded (using update utilities).
Mode Select | SW1 | SW2 | SW3 | SW4 | SW5 | SW6 | SW7 | SW8 |
---|---|---|---|---|---|---|---|---|
TEST/BOOT | 1 | 1 |
1 |
0 |
0 |
0 |
0 |
0 |
0 (switch down), 1 (switch up), SW (switch)
A forced cold start can be used to bring the module into a defined mode with default parameter settings.
CAUTION
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions are calibration settings, diagnosis and repair log books which will not be erased.
If you use the following switch settings and power the instrument up again, a forced cold start has been completed.
Mode Select | SW1 | SW2 | SW3 | SW4 | SW5 | SW6 | SW7 | SW8 |
---|---|---|---|---|---|---|---|---|
TEST/BOOT | 1 | 1 |
0 |
0 |
0 |
0 |
0 |
1 |
0 (switch down), 1 (switch up), SW (switch)