Error ID: 29227
Power consumption too high for one of the MCGV valves.
Probable cause | Suggested actions | |
---|---|---|
1 | Cable defect. | Replace MCGV/4CGV. |
2 | Valve defect. | Please contact your Agilent service representative. Replace MCGV/4CGV. |
3 | Defective mainboard. | Please contact your Agilent service representative. |
base-id: 4133123467
id: 4133123467