Error ID: 29181
A target condition (run volume, which is the volume delivered since the method run start) has been sent to the instrument which should have been reached within an expected time frame but didn’t. Either the limit, time frame or the current value of the variable has been modified later directly or indirectly (for example the flow rate).
Probable cause | Suggested actions | |
---|---|---|
1 | Incorrect parameters have been sent to the instrument by the control software or manual changes. | Verify control software, macros, manual commands. |
base-id: 4132977803
id: 4132977803