For LC Drivers 3.4 or newer, please use Ammonium Sulfate in Water V.04 and Sodium Chloride in Water V.04 and manually input the molarity (0.5 M - 3.0 M or 0.5 M - 2.0 M respectively) as shown in the example below.
With the 1290 Infinity III Bio LC System, Agilent offers a dedicated instrument for running applications that are typical for the (Bio)pharma market, such as IEX (Ion Exchange Chromatography) or HIC (Hydrophobic Interaction Chromatography), and applications in Academia, Government and Applied Markets, using concentrated salt solutions as eluents.
For LC Drivers 3.3 or older, please choose the solvent definition most suitable for the application according to the chart below.
NOTE
Ammonium Sulphate V.03 and Sodium Chloride V.03 only properly characterize the G7132A High-Speed Pump. Do not use these with any quaternary/flexible pump configurations.
Main salt component | Total salt content (c, mol/L) | Select solvent definition |
---|---|---|
NaCl | 0.25 < c ≤ 0.75 | Sodium Chloride 0.5 M V.03 |
0.75 < c ≤ 1.25 | Sodium Chloride 1.0 M V.03 | |
1.25 < c ≤ 1.75 | Sodium Chloride 1.5 M V.03 | |
c > 1.75 | Sodium Chloride 2.0 M V.03 | |
(NH4)₂SO4 | 0.25 < c ≤ 0.75 | Ammonium Sulphate 0.5 M V.03 |
0.75 < c ≤ 1.25 | Ammonium Sulphate 1.0 M V.03 | |
1.25 < c ≤ 1.75 | Ammonium Sulphate 1.5 M V.03 | |
1.75 < c ≤ 2.5 | Ammonium Sulphate 2.0 M V.03 | |
c >2.5 | Ammonium Sulphate 3.0 M V.03 |