The pump offers a powerful feature to monitor the liquid level in the solvent bottles. With total bottle volume and initial filling volume set correctly, the pump subtracts the consumed volume continuously from the initial value and stops the pump and method/sequence execution before the system runs dry or an analysis is corrupted.
CAUTION
The bottle filling feature fails if multiple solvent inlets are put into one solvent bottle!
Bottle Filling Parameters lists the available bottle filling parameters.
Parameter | Limits | Description |
---|---|---|
| 0 – 1000 L default: 0 L | This is the capacity (maximum possible volume) in liter of the solvent bottle. In combination with the actual volume, this parameter is used for calculating and displaying the relative liquid level. |
| 0 – 1000 L default: 0 L | After filling the solvent bottles, enter the actual volumes into these boxes. The Actual Volume must not be larger than the Total Volume of the bottle. |
| default: unchecked | If this option is checked, the pump won’t start a new run if the solvent level in one or more bottles is below the minimum volume. Enter a minimum volume in liter, which considers the position of the solvent inlet and size/shape of the solvent bottle such that no air is drawn if the actual volume gets close to this limit. |
| default: unchecked | If this option is checked, the pump will turn off before air is aspirated. However, the residual solvent volume has been calculated for 1 L solvent bottles and may be too small for large bottles or other vessels. |