A leak of the outlet valve will be identified separately (internal outlet valve leak) by calculating the difference between leak rate 1 and leak rate 2. If the second leak rate is higher than the first one, this is due to a flow back through the outlet valve.
Probable cause | Suggested actions | |
---|---|---|
1 | Leak at outlet valve | Replace the part which has failed and re-run the test. |
base-id: 3820731787
id: 3820731787-1