Error ID: 6707, 6708
During initialization of the detector, the excitation- and emission monochromator are activated and the encoder should generate an index.
Probable cause | Suggested actions | |
---|---|---|
1 | Encoder defective. | Exchange the monochromator assembly. |
2 | Encoder electric defective. | Replace FLM board. |
3 | Monochromator defective or missing. | Check for connection to FLF (Ex) and FLM (Em) or replace monochromator assembly. |
4 | One phase monochromator motor power driver defective. | Defective FLF board (Ex) or FLM board (Em). |
base-id: 3604027019