Error ID: 6705, 6706
During initialization of the detector, the excitation and emission monochromator are activated.
Probable cause | Suggested actions | |
---|---|---|
1 | Encoder cables mixed on FLM board. | Check encoder connections to FLM. |
2 | Monochromator assembly not connected. | Check motor connections to FLF (Ex) and FLM (Em) and encoders to FLM. |
3 | Monochromator or encoder defective. | Exchange the monochromator assembly. |
4 | Monochromator motor power driver defective. | Defective FLF board (Ex) or FLM board (Em). |
base-id: 3603926667