Type | Specification | Comments |
---|---|---|
Weight | 17.8 kg (39.2 lbs) |
|
Dimensions (height × width × depth) | 180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches) |
|
Line voltage | 100 – 240 V~, ± 10 % | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ± 5 % |
|
Power consumption | 90 VA / 74 W |
|
Ambient operating temperature | 4 – 55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40 – 70 °C (-40 – 158 °F) |
|
Humidity | < 95 % r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Installation category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Hydraulic system | Two dual piston in series pumps with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons |
|
Designed for use with Agilent InfinityLab Assist Technology | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting | |
Flow range | Settable: 0.001 – 5 mL/min | Recommended: 1.0 – 5.0 mL/min (in 0.001 mL/min increments) |
Flow precision | ≤ 0.07 % RSD or < 0.02 min SD, whichever is greater | Based on retention time at constant temperature |
Pressure operating range | Up to 60 MPa (600 bar, 8702 psi) up to 5 mL/min |
|
Compressibility compensation | Pre-defined, based on mobile phase compressibility |
|
Recommended pH range | 1.0 – 12.5 | Solvents with pH < 2.3 should not contain acids that attack stainless steel |
Gradient formation | High-pressure binary mixing |
|
Delay volume | Standard delay volume configuration: 600 – 900 µL, (includes 400 µL mixer), dependent on back pressure |
|
Composition range | settable: 0 – 100 % |
|
Composition precision | < 0.15 % RSD or < 0.04 min SD, whichever is greater | Based on retention time at constant temperature |
Integrated degassing unit | Number of channels: 2 Internal volume per channel: 1.5 mL |
|
Instrument Control | LC & CE Drivers A.02.16 or above Instrument Control Framework (ICF) A.02.04 or above Instant Pilot (G4208A) with firmware B.02.21 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Lab Advisor software B.02.09 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers |
Communication | Controller-area network (CAN), Extended Remote Interface (ERI), Local Area Network (LAN) |
|
Safety features and maintenance | Leak detection, safe leak handling, leak output signal for shutdown of the pumping system. Low voltage in major maintenance areas. Extensive diagnostics, error detection and display with Agilent Lab Advisor software. |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors. |
|
Housing | All materials are recyclable |
|
base-id: 9597896459