Type | Specification | Comments |
---|---|---|
Weight | 11.9 kg (26.2 lbs) |
|
Dimensions | 140 x 396 x 436 mm |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 70 VA, 60 W |
|
Ambient operating temperature | 4–40 °C (39–104 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Detection type | One signal wavelength (excitation and emission) | Programmable single wavelength (excitation and emission) fluorescence detector |
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting | |
Single wavelength operation |
|
|
Light source | Xenon Flash Lamp, normal mode 20 W, economy mode 5 W, lifetime 4000 h |
|
Pulse frequency | 296 Hz for single signal mode 74 Hz for economy mode |
|
Maximum data rate | 74 Hz |
|
Excitation monochromator | Range: settable 200 nm - 1200 nm and zero-order Bandwidth: 20 nm (fixed) |
|
Emission monochromator | Range: settable 200 nm - 1200 nm and zero-order Bandwidth: 20 nm (fixed) |
|
Reference system | In-line excitation measurement |
|
Timetable programming | Single signal wavelength, response time, PMT Gain, baseline behavior (append, free, zero) |
|
Wavelength characteristic | Repeatability +/- 0.2 nm Accuracy +/- 3 nm setting |
|
Flow cells | Standard: 8 µL volume and 20 bar (2 MPa) pressure maximum, fused silica block |
|
Analog output | Recorder/integrator: 100 mV or 1 V, output range > 100 LU, two outputs | 100 LU is the recommended range |
Instrument Control | LC & CE Drivers A.02.14 or above Instrument Control Framework (ICF) A.02.03 or above Lab Advisor software B.02.09 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Instant Pilot (G4208A) with firmware B.02.19 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers |
Communications | Controller Area Network (CAN), USB, ERI: ready, start, stop and shut-down signals |
|
Safety features and maintenance | Leak detection, safe leak handling, leak output signal for shutdown of the pumping system. No hazardous voltages in major maintenance areas. Extensive diagnostics, error detection and display with Agilent InfinityLab Assist and Agilent Lab Advisor software. |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors. |
|
Housing | All materials are recyclable. |
|
base-id: 6097251211
id: 36028803116215179