Parameter Name | External Value | Comment |
---|---|---|
Weight | 12 kg (26.5 lbs) |
|
Dimensions (height × width × depth) | 140 x 396 x 436 mm (5.5 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10 % | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5 % |
|
Power consumption | 110 VA, 100 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95 % r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification |
---|---|
Detection type | 1024-element photodiode array |
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting |
Light source | Deuterium and tungsten lamps |
Number of signals | 8 |
Maximum data rate | 120 Hz |
Short term signal noise (ASTM) | < ± 0.7·10-5 AU at 254 and 750 nm |
Drift | < 0.9·10-3 AU/h at 254 nm and 750 nm |
Linear absorbance range | > 2 AU (5 %) at 273 nm |
Wavelength range | 190 – 950 nm |
Wavelength accuracy | ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter |
Wavelength bunching | 1 – 400 nm, programmable in 1 nm steps |
Slit width | 1, 2, 4 , 8, 16 nm |
Diode width | ~ 1 nm |
Time programmable | Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength range, threshold, spectra storage mode |
Flow cells | Standard: 13 µL volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum Standard bio-inert: 13 µL volume, 10 mm cell path length and 120 bar (1740 psi) pressure maximum Semi-micro: 5 µL volume, 6 mm cell path length and 120 bar (1740 psi) pressure maximum Micro: 2 µL volume, 3 mm cell path length, 120 bar (1740 psi) pressure maximum Semi-nano: 500 nL volume, 10 mm cell path length and 40 bar (580 psi) pressure maximum Nano: 80 nL volume, 6 mm cell path length and 40 bar (580 psi) pressure maximum High pressure: 1.7 µL volume, 6 mm cell path length and 400 bar (5800 psi) pressure maximum Prep SST: 3 mm cell path length and 120 bar (1740 psi) pressure maximum Prep Quartz: 3 mm cell path length and 50 bar (1740 psi) pressure maximum Prep Quartz:0.3 mm cell path length and 50 bar (1740 psi) pressure maximum Prep Quartz: 0.06 mm cell path length and 50 bar (1740 psi) pressure maximum SFC Flow Cell: 13 μL volume, 10 mm cell path length and 400 bar (5800 psi) pressure maximum SFC Flow Cell LD: 2 μL volume, 3 mm cell path length and 400 bar (5800 psi) pressure maximum |
Spectral tools | Data analysis software for spectra evaluation, including spectral libraries and peak purity functions |
Analog output | Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output |
Instrument control | LC and CE Drivers A.02.14 or above Instrument Control Framework (ICF) A.02.04 or above Lab Advisor B.02.08 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Agilent Instant Pilot (G4208A) with firmware B.02.20 or above For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers. |
Communication | LAN, Controller-Area Network (CAN), USB Extended Remote Interface (ERI): ready, start, stop and shut-down signals |
GLP | RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage) Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-setable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter. |
Safety and maintenance | Extensive diagnostics, error detection and display through Agilent Instant Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
Housing | All materials recyclable. |
Others | Second generation of Electronic temperature control (ETC) for the complete optical unit |
base-id: 6079254795
id: 36028803098218763