Type | Specification | Comments |
---|---|---|
Weight | 16.1 kg (35.5 lbs) |
|
Dimensions (height × width × depth) | 180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 120 VA / 110 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Hydraulic system | Dual pistons in series pump with proprietary servo-controlled variable stroke design and smooth motion control for active damping. |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, | |
Pump resolution step size | 300 pL |
|
Flow range | Settable: 0.001 – 5 mL/min | In 0.001 mL/min increments |
Flow precision | ≤ 0.07 % RSD or 0.01 min SD, whichever is greater |
|
Flow accuracy | ± 1 % or ± 10 µL/min, whichever is greater | Pumping degassed H₂O |
Pressure operating range | Up to 130 MPa (1300 bar, 18855 psi) up to 2 mL/min Ramping down to 80 MPa (800 bar, 11603 psi) at 5 mL/min |
|
Pressure pulsation | < 1 % amplitude or < 0.5 MPa (5 bar), whichever is greater |
|
Compressibility compensation | Automatic | When using "Solvent Types" in method |
Recommended pH-range | 1.0 – 12.5 | Solvents with pH < 2.3 should not contain acids which attack stainless steel. |
Gradient formation | Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve |
|
Delay volume | ≤ 350 µL (default configuration) | Measured with water at 1 mL/min (water/water with tracer) |
Composition range | Settable: 0 – 100 % Recommended: 1 – 99 % |
|
Composition precision | < 0.15 % RSD or 0.02 min SD, whichever is greater |
|
Composition accuracy | ± 0.4 % absolute | At 1 mL/min for water/water with tracer |
Number of solvents | 4 |
|
Solvent selection valve | Internal 4-solvent gradient formation valve included. External 2x 12 solvent valve as option, fully integrated in the pump control interface. |
|
Integrated degassing unit | Number of channels: 4 Internal volume per channel: 1.5 mL |
|
Materials in contact with solvent | TFE/PDD copolymer, PFA, PEEK, FEP, ETFE, Stainless Steel, Aluminum Oxide-based Ceramic, Ruby, Sapphire, Gold, Zirconium Oxide-based Ceramic, Tantalum, UHMW-PE, Polyimide, DLC Coated Stainless Steel |
|
Automatic purge valve | Enables automatic software-embedded functionalities such as switching the optional mixer in and out or automatic purging. |
|
Active Seal wash | Included |
|
Intelligent System Emulation Technology (ISET) | Included |
|
Instrument control | LC & CE Drivers A.02.11 or above Instrument Control Framework (ICF) A.02.03 or above Lab Advisor software B.02.06 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Instant Pilot (G4208A) with firmware B.02.17 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC & CE Drivers. |
Communication | Controller Area Network (CAN), LAN, RS232C, APG remote: ready, start, stop, and shutdown signals |
|
Safety features and maintenance | Leak detection, safe leak handling, leak output signal for shutdown of the pumping system. No hazardous voltages in major maintenance areas. Extensive diagnostics, error detection and display with Agilent Lab Advisor software. |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors. |
|
Housing | All materials are recyclable. |
|
|
| Performance specifications determined with standard configuration (4CGV). |