For bio-inert modules use bio-inert parts only! |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 5023-2524 | Hex Key Set (part of the G7120-62708 InfinityLab LC Series Tool Kit) |
1 | ![]() | 01018-23702 | Insert tool (for 100 µL seals) OR |
1 | ![]() | G4226-43800 | Seal insert tool (for 40 µL seals) |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 0905-1717 | Metering Seal, 40 µL |
1 | ![]() | 0905-1719 | PE Seal |
1 | ![]() | G5611-21503 | Seal PTFE (Bio-inert) for bio inert solution |
1 | ![]() | G7137-20003 | Metering seal 1290 Bio 2 mm piston, 40 µL |
1 | ![]() | 0905-1294 | Metering Seal, 900 µL |
1 | ![]() | 5067-5920 | Piston, 40 µL, Zirconium oxide If previous piston is scratched (for 40 µL analytical head) |
1 | ![]() | 5067-5678 | Piston, 100 µL, Zirconium oxide If previous piston is scratched (for 100 µL analytical head) |
1 | ![]() | G4267-60462 | Piston, 900 µL, Sapphire If previous piston is scratched (for 900 µL analytical head) |
In the Local Controller start the maintenance mode and select Service Position.
OR
In the Agilent Lab Advisor software select Service & Diagnostics in the system screen (Tools) > Maintenance Positions > Change Metering Device , click start and wait until the metering device is in maintenance position.
Take the metering device. Push against the rear side of the metering device and rotate a quarter left to release the bayonet lock.
If dirty: Clean the piston with an appropriate solvent. If scratched: Replace the piston by a new one.
Carefully remove the metering seal using the steel side of the insert tool. Clean the chamber with an appropriate solvent and ensure that all particulate matter is removed.
base-id: 3802848011