Error ID: 63
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals. A LOW signal input on pin 4 of the remote connector generates the error message.
The module continually monitors the remote input connectors for status signals. A LOW signal input on pin 6 of the Enhanced Remote Interface (ERI) connector generates the error message.
Probable cause | Suggested actions | |
---|---|---|
1 | Leak detected in another module with a CAN connection to the system. | Fix the leak in the external instrument before restarting the module. |
2 | Leak detected in an external instrument with a remote connection to the system. | Fix the leak in the external instrument before restarting the module. |
3 | Shut-down in an external instrument with a remote connection to the system. | Check external instruments for a shut-down condition. |
4 | The degasser failed to generate sufficient vacuum for solvent degassing. | Check the external vacuum degasser module (if installed) for an error condition. Refer to the Service Manual for the degasser or the pump that has the degasser built-in. |