Error ID: 2686
The electronic fuse on the purge valve has been activated.
Probable cause | Suggested actions | |
---|---|---|
1 | Short in purge valve circuit. | Power cycle the module. |
2 | Defective purge valve. | Please contact your Agilent service representative. |
3 | Defective mainboard. | Please contact your Agilent service representative. |
base-id: 3596704523