The cuvette is used for off-line measurements (no flow system required) and is basically a standard flow cell with a few changes:
wide bore capillary connections for easier injections with a syringe
identification lever for cell auto-recognition system.
Install the cuvette instead of the standard flow cell.
Connect the waste tubing to the outlet of the cuvette.
Use the syringe (see Cuvette Kit) to inject the compound.
Setup the parameters for the Fluorescence Scan (under Special Setpoints).
Select “Take Fluorescence Scan” on the user-interface to start the off-line measurement.
base-id: 3586960267
id: 3586960267