Error ID: 6721
Revision A/B/C Front End Board (FLF):
There is no feedback mechanism that checks whether the lamp is ON! If no peaks are shown in the chromatogram, the user-interface shows the module still in Ready. Perform a "Lamp Intensity Test" (see Lamp Intensity Test) first. If flat use below steps.
Revision A/B/C Front End Board (FLF):
There is no feedback mechanism that checks whether the lamp is ON! If no peaks are shown in the chromatogram, the user-interface shows the module still in Ready. Perform a "Lamp Intensity Test" (see --- MISSING LINK ---) first. If flat use below steps.
Revision D Front End Board (FLF):
The flashing of the xenon flash lamp is monitored constantly. If the Lamp has not flashed for more than 100 times in series, an error is generated and the lamp is turned OFF.
Probable cause | Suggested actions | |
---|---|---|
1 | Defective Hardware. | Please contact your Agilent service representative. |