To achieve optimum limits of detection and selectivity, analysts must find out about the fluorescent properties of the compounds of interest. Excitation and emission wavelengths can be selected for optimum limits of detection and best selectivity. In general, fluorescence spectra obtained with different instruments may show significant differences depending on the hardware and software used.
The traditional approach is to extract an appropriate excitation wavelength from the UV spectrum that is similar to the fluorescence excitation spectrum (see Excitation and emission spectra of quinidine) and to record the emission spectrum. Then with an optimum emission wavelength determined, the excitation spectrum is acquired.
These tasks have to be repeated for each compound using either a fluorescence spectrophotometer or stop-flow conditions in LC. Usually each compound requires a separate run. As a result, a set of excitation and emission spectrum is obtained (Isofluorescence plot of a mobile phase) for each compound. Since this is a tedious procedure, it is applicable only when there is a limited number of compounds of interest.
The Agilent 1200 Infinity Series LC offers three different ways to obtain complete information on a compound's fluorescence:
Procedure I - Take a fluorescence scan offline for a single compound as described above for the mobile phase. This is done preferably with a manual FLD cuvette when pure compounds are available.
Procedure II - Use two LC runs with the Agilent 1260 Infinity Fluorescence Detector to separate the compound mix under known conditions and acquire emission and excitation spectra separately.
Procedure III - Use an Agilent 1200 Infinty Series FLD/DAD combination and acquire UV/Visible spectra (equivalent to excitation spectra) with the DAD and emission spectra with the FLD-both in a single run.
Because fluorescence spectra traditionally have not been easily available with previous LC fluorescence detectors, standard fluorescence spectrophotometers have been used in the past to acquire spectral information for unknown compounds. Unfortunately this approach limits optimization, as there are differences expected in optical design between an LC detector and a dedicated fluorescence spectrophotometer, or even between detectors. These differences can lead to variations for the optimum excitation and emission wavelengths.
The Agilent 1260 Infinity Fluorescence Detector offers a fluorescence scan that delivers all spectral information previously obtained with a standard fluorescence spectrophotometer, independent of the LC fluorescence detector. Characterization of a pure compound from a fluorescence scan shows the complete information for quinidine as obtained with the Agilent 1260 Infinity Fluorescence Detector and a manual cuvette in a single offline measurement. The optima for excitation and emission wavelengths can be extracted as coordinates of the maxima in the three dimensional plot. One of the three maxima in the center of the plot can be chosen to define the excitation wavelength. The selection depends on the additional compounds that are going to be analyzed in the chromatographic run and the background noise that may be different upon excitation at 250 nm, 315 nm or 350 nm. The maximum of emission is observed at 440 nm.
Details for Characterization of a pure compound from a fluorescence scan:
All excitation and emission spectra of Quinidine (1 µg/ml) are shown in graphic. Fluorescence intensity is plotted vs excitation and emission wavelengths.
Detector settings: step size 5 nm, PMT 12 , Response time 4 s
The conditions for the separation of organic compounds such as polyaromatic nuclear hydrocarbons (PNAs) are well described in various standard methods, including commonly used EPA and DIN methods. Achieving the best detection levels requires checking for the optimum excitation and emission wavelengths for all compounds. Yet taking fluorescence scans individually makes this a tedious process. A better approach is to acquire spectra online for all compounds during a run. This speeds up method development tremendously. Two runs are sufficient for optimization.
During the first run, one wavelength is chosen in the low UV range for the excitation wavelength and one emission wavelength in the spectral range for the emission wavelength. Most fluorophores show strong absorption at these wavelengths and the quantum yield is high. Excitation is sufficient for collecting emission spectra.
Optimization of the time-program for the emission wavelength contains all emission spectra obtained in a single run from a mix of 15 PNAs. This set of spectra is used to set up a timetable for optimum emission wavelengths for all compounds.
The individual compound spectra in the isofluorescence plot show that at least three emission wavelengths are needed to detect all 15 PNAs properly:
0 min: |
350 nm |
for naphthalene to phenanthrene |
8.2 min: |
420 nm |
for anthracene to benzo(g,h,i)perylene |
19.0 min: |
500 nm |
for indeno(1,2,3-c,d)pyrene |
In the second run, three setpoints for emission wavelengths are entered into the time-program and excitation spectra are recorded, as shown in Optimization of the time-program for the excitation wavelength. The area of high intensity (red) is caused by stray light when emission spectra overlap with the excitation wavelength. This can be avoided by fitting the spectral range automatically. Excitation at 260 nm is most appropriate for all PNAs.
Column |
Vydac, 2.1 x 200 mm, PNA, 5 µm |
Mobile phase |
A = water; B = acetonitrile (50 : 50) |
Gradient |
3 minutes, 60% 14 minutes, 90% 22 minutes, 100% |
Flow rate |
0.4 ml/min |
Column temperature |
18 °C |
Injection volume |
5 µl |
FLD settings |
PMT 12, response time 4 s, step size 5 nm |
The obtained data are combined to setup the time-table for the excitation wavelength for best limit of detection and selectivity. The optimized switching events for this example are summarized in Timetable for the analysis of 15 polynuclear aromatic hydrocarbons.
Time [min] | Exitation Wavelength [nm] | Emission Wavelength [nm] |
---|---|---|
0 |
260 |
350 |
8.2 |
260 |
420 |
19.0 |
260 |
500 |
This timetable gives the conditions for optimum detection based on the results of two chromatographic runs.
For most organic compounds, UV-spectra from diode array detectors are nearly identical to fluorescence excitation spectra. Spectral differences are caused by specific detector characteristics such as spectral resolution or light sources.
In practice, combining a diode array detector with a fluorescence detector in series gives the full data set needed to achieve the optimum fluorescence excitation and emission wavelengths for a series of compounds in a single run. With the UV/Visible/excitation spectra available from the diode array detector, the fluorescence detector is set to acquire emission spectra with a fixed excitation wavelength in the low UV range.
The example is taken from the quality control of carbamates. Samples are analyzed for the impurities 2,3-diaminophenazine (DAP) and 2-amino-3-hydroxyphenazine (AHP). Reference samples of DAP and AHP were analyzed with diode array and fluorescence detection. UV-spectrum and fluorescence spectra for 2,3-diaminophenazine (DAP) shows the spectra obtained from both detectors for DAP. The excitation spectrum of DAP is very similar to the UV absorption spectrum from the diode array detector. Qualitive analysis of MBC (2-benzimidazole carbamic acid methylester) and impurities shows the successful application of the method to a carbamate sample and a pure mixture of DAP and AHP for reference. The column was overloaded with the non-fluorescent carbamate (2-benzimidazole carbamic acid methylester/MBC) to see the known impurities, AHP and DAP.
Column |
Zorbax SB, 2 x 50 mm, PNA, 5 µm |
Mobile phase |
A = water; B = acetonitrile |
Gradient |
0 minutes, 5% 10 minutes, 15% |
Flow rate |
0.4 ml/min |
Column temperature |
35 °C |
Injection volume |
5 µl |
FLD settings |
PMT 12, response time 4 s, step size 5 nm Ex 265 nm and 430 nm Em 540 nm |
base-id: 3586005771
id: 3586005771