The example below uses water.
Pump solvent through your system.
Set the fluorescence scan range under FLD special setpoints according to your needs.
NOTE
The scan time will increase when the range is enlarged. With the default values, the scan takes about 2 minutes.
Set PMT gain to 16.
Define a data file name and take a fluorescence scan. After the scan is completed, the isoabsorbance scan results appear, see Fluorescence Scan of Water.
NOTE
A low background will improve the signal-to-noise, see also Reducing Stray Light.