This example uses the following chromatographic conditions (the detector settings are shown in Detector settings for emission scan).
Mobile phases |
A = water = 50 % B = Acetonitrile = 50 % |
Column |
Vydac-C18-PNA, 250 mm x 2.1 mm i.d. with 5 µm particles |
Sample |
PAH 0.5 ng |
Flow rate |
0.4 ml/min |
Compressibility A (water) |
46 |
Compressibility B (Acetonitrile) |
115 |
Stroke A and B |
auto |
Time Table |
at 0 min % B=50 |
|
at 3 min % B=60 |
|
at 14.5 min % B=90 |
|
at 22.5 min % B=95 |
Stop time |
26 min |
Post time |
8 min |
Injection volume |
1 µl |
Oven temperature (1200) |
30 °C |
FLD PMT Gain |
PMT = 15 |
FLD Response time |
4 s |
Wait until the baseline stabilizes. Complete the run.
Load the signal. (In this example just the time range of 13 min is displayed).
Use the isoabsorbance plot and evaluate the optimal emission wavelengths, shown in the table below.
Peak # | Time | Emission Wavelength |
---|---|---|
1 | 5.3 min | 330 nm |
2 | 7.2 min | 330 nm |
3 | 7.6 min | 310 nm |
4 | 8.6 min | 360 nm |
5 | 10.6 min | 445 nm |
6 | 11.23 min | 385 nm |
Using the settings and the timetable (from previous page), do a second run for the evaluation of the optimal excitation wavelength.
Wait until the baseline stabilizes. Start the run.
Load the signal.
Use the isoabsorbance plot and evaluate the optimal excitation wavelengths (in this example just in the time range of 13 minutes).
The table below shows the complete information about emission (from Use the isoabsorbance plot and evaluate the optimal emission wavelengths, shown in the table below.) and excitation maxima.
Peak # | Time | Emission Wavelength | Excitation Wavelength |
---|---|---|---|
1 | 5.3 min | 330 nm | 220 / 280 nm |
2 | 7.3 min | 330 nm | 225 / 285 nm |
3 | 7.7 min | 310 nm | 265 nm |
4 | 8.5 min | 360 nm | 245 nm |
5 | 10.7 min | 445 nm | 280 nm |
6 | 11.3 min | 385 nm | 270 / 330 nm |
base-id: 3585981195
id: 3585981195