The lamp flash frequency can be changed into the following modes:
Positioning |
296 Hz (Standard), 560 V |
63 mJ (18.8 W) |
|
74 Hz (Economy), 560 V |
63 mJ (4.7 W) |
Rotation (Multi Ex/Em) |
74 Hz (Standard), 950 V |
180 mJ (13.3 W) |
|
74 Hz (Economy), 560 V |
63 mJ (4.7 W) |
Best sensitivity can be expected with “no economy”, see Xenon Flash Lamp Frequency.
base-id: 3585931659