Perform the following steps to diagnose whether the lamp is the cause of the baseline instability:
Ensure that the detector has been properly warmed up, see Overview of Diagnostic Signals.
Flush the flow cell with water and start runs without sample injections (blank runs).
Use a restriction capillary to remove any influence of the column.
If you see instabilities, stop the flow and compare.