The detector has several signals (internal temperatures, Reference only) that can be used for diagnosing problems. These can be
baseline problems deriving from lamp,
wander / drift problems due to temperature changes.
For intermittent baseline problems the board temperature should be monitored to see the impact of it to the signal.
NOTE
Measurements should be taken over 60 – 120 min minimum to include long-term effects (e.g. cycle of air condition systems).