Error ID:
NOTE
If the test fails with Max-Light Cartridge Test Cell and new UV-lamp, the optical unit must be replaced.
Probable cause | Suggested actions | |
---|---|---|
1 | Absorbing solvent or air bubble in Max-Light Cartridge Cell. | Repeat calibration with Max-Light Cartridge Test Cell and compare results. |
2 | Dirty or contaminated Max-Light Cartridge Cell. | Ensure the Max-Light Cartridge Cell is filled with water. Recalibrate. |
3 | Old UV-lamp. | Exchange the UV lamp. |
4 | Dirty or contaminated optical components. | Run the Cell Test. If the test fails, flush the flow cell. See also Clean the Max-Light Cartridge Cell. |