The cell test measures the intensity of the UV-lamp over the full wavelength range (190 - 690 nm), once with the Max-Light Cartridge Cell installed, and once with the Max-Light Cartridge Test Cell. The resulting intensity ratio is a measure of the amount of light absorbed by the Max-Light Cartridge flow cell. The test can be used to check for dirty or contaminated flow cell windows. When the test is started, the 1-nm slit is moved into the light path automatically (G7117B only). On the G7117A/G7117C, the 4 nm fixed slit is used.
This test should be performed initially with a new detector/flow cell. The values should be kept for later reference/comparison.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Max-Light Cartridge Cell (filled with water) |
1 | ![]()
|
| Max-Light Cartridge Test Cell |
Prerequisites
|
Run the Cell-Test with Agilent Lab Advisor (for further information see Online-Help of user interface).