The ASTM noise test determines the detector noise over a period of 20 minutes. The test is done with installed Max-Light Cartridge Cell or Max-Light Cartridge Test Cell.
This test does also check for the drift. It is also part of the “Self Test” (without checking for the drift).
If the test is performed with the Max-Light Cartridge Test Cell, the test results are not influenced by solvent or pump effects.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Max-Light Cartridge Cell (filled with water) |
1 | ![]()
|
| Max-Light Cartridge Test Cell |
Prerequisites
|
Run the ASTM Drift and Noise Test with Agilent Lab Advisor (for further information see Online-Help of user interface).
base-id: 3526336907