The quick noise test measures the noise of the detector, with Max-Light Cartridge Cell or with Max-Light Cartridge Test Cell installed, in one minute intervals over a total of 5 minutes.
The noise of the detector is calculated by using the maximum amplitude for all random variations of the detector signal of frequencies greater than one cycle per hour. The noise is determined for 5 one minute intervals and is based on the accumulated peak-to-peak noise for the intervals. At least seven data points per cycles are used in the calculation. The cycles in the noise determination are not overlapping.
If the test is performed with the Max-Light Cartridge Test Cell, the test results are not influenced by solvent or pump effects.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Max-Light Cartridge Cell (filled with water) |
1 | ![]()
|
| Max-Light Cartridge Test Cell |
Prerequisites
|
Run the Quick Noise Test with Agilent Lab Advisor (for further information see Online-Help of user interface).