| The figure shows the default settings.
Lamps: can be turned ON/OFF. At Power On: automatic lamp-on at power on. Analog Output Range: can be set to either 100 mV or 1 V full scale, for additional settings see Analog Output (under Method Parameter Settings). Temperature Control: can be turned ON/OFF. The optical unit is kept on constant temperature and improves the baseline stability in unstable environments. See also note below. ON it will keep the optical unit stabilized. UV Lamp Tag Use lamp only if lamp tag is available detects a lamp with RFID tag. If no RFID tag lamp is used, “UV lamp not ready” is displayed and it cannot be ignited. A compatible mode has to be selected based on the used lamp; see Non-RFID-tag lamp information below. Always use lamp: In case a non-RFID-tag lamp is used, the user interface will show this when selecting a compatible mode. You may operate the detector outside of the guaranteed specification. The correct selection is important for optimal performance and lifetime.
Cell Tag Use cell only if cell tag is available detects a cell with RFID tag. If no RFID tag cell is used, “cell tag not ready” is displayed and it cannot be ignited and analysis is disabled. Always use cell: In case a non-RFID-tag cell is used, the user interface will show this when selecting a compatible mode.
Automatic Turn On: automatic detector power on.
|