Test evaluation
NOTE
If the test fails with Flow Cell and new UV-lamp, the optical unit must be checked/replaced.
Probable cause | Suggested actions | |
---|---|---|
1 | Absorbing solvent or air bubble in Flow Cell. | Repeat calibration with Flow Cell and compare results. |
2 | Dirty or contaminated Flow Cell. | Ensure the flow cell is filled with water, and free from air bubbles. Recalibrate. |
3 | Old UV-lamp. | Exchange the UV lamp. |
4 | Dirty or contaminated optical components. | Run the Cell Test. If the test fails, flush the flow cell. |