The ASTM noise test determines the detector noise over a period of 20 minutes. The test is done with installed flow cell or flow cell removed.
This test also checks for the drift. It is also part of the Self Test (without checking for the drift).
If the test is performed with the flow cell removed, the test results are not influenced by solvent or pump effects.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Removed Flow Cell |
Prerequisites
|
Run the ASTM Drift and Noise Test with Agilent Lab Advisor (for further information see Online-Help of user interface).
base-id: 3358701451
id: 3358701451