All tests are described based on the Agilent Lab Advisor Software B.02.08. Other user interfaces may not provide any test or just a few.
Interface | Comment | Available Function |
---|---|---|
Agilent Lab Advisor | For functions, refer to Function Overview Lab Advisor | Available functions depend on Product Level (Basic – Advanced – FSE) |
Agilent ChemStation | No tests available Adding of temperature/lamp signals to chromatographic signals possible |
|
Function | Product Level | |
---|---|---|
Tests |
|
|
- ASTM Drift and Noise Test | Basic | Advanced |
- Cell Test | Basic | Advanced |
- D/A Converter Test | Basic | Advanced |
- Dark Current Test | Basic | Advanced |
- Filter/Grating Motor Test | Basic | Advanced |
- Holmium Oxide Test | Basic | Advanced |
- Intensity Test | Basic | Advanced |
- Quick Noise Test | Basic | Advanced |
Calibrations |
|
|
- Wavelength Calibration | Basic | Advanced |
Tools |
|
|
- Diagnostic Buffers | Basic | Advanced |
- Board Check and Change |
|
|
- Module Info | Basic | Advanced |
- Firmware Declustering |
|
|
- Test Chromatogram | Basic | Advanced |
- Spectral Scan | Basic | Advanced |
Controls |
|
|
- Advanced Method Parameters |
|
|
- Analog Output 1 Attenuation |
| Advanced |
- D2 lamp required |
| Advanced |
- Analog Output 1 Offset [% Full Scale] |
| Advanced |
- Configuration |
|
|
- Remote Pulse Duration [s] * | Basic | Advanced |
- Analog Output 1 Range |
| Advanced |
- Control |
|
|
- Balance Detector |
| Advanced |
- UV Lamp | Basic | Advanced |
- Conversions |
|
|
- Transfer to Resident FW |
|
|
- Transfer to Main FW * |
|
|
- G7114B allows G1314E and G1314F | Basic | Advanced |
- G7114A allows NONE | Basic | Advanced |
- Method Parameters |
|
|
- Set Data Rate [Hz] |
| Advanced |
- Set Wavelength [nm] |
| Advanced |
- Module Information |
|
|
- Firmware Version |
|
|
- Identify Module | Basic | Advanced |
- Special Commands |
|
|
- Lamp tag required | Basic | Advanced |
- Cell tag required | Basic | Advanced |
- Clear Error | Basic | Advanced |
- Detector Reset | Basic | Advanced |
- Forced Cold Start |
|
|
Actuals |
|
|
- Signal A [mAU] |
| Advanced |
- Sample Signal |
|
|
- Reference Signal |
|
|
Statemachines |
|
|
- UV Lamp | Basic | Advanced |
Signals |
|
|
- Signal A [mAU] |
| Advanced |
- Sample Signal [mAU] |
| Advanced |
- Reference Signal [mAU] |
| Advanced |
- Board Temperature [°C] |
| Advanced |
- Lamp Voltage [V] |
| Advanced |
EMF Counters |
|
|
- Accumulated UV Lamp On-Time | Basic | Advanced |
- Number of UV Lamp Ignitions | Basic | Advanced |