Probable cause | Suggested actions | |
---|---|---|
1 | Detector not calibrated. | Recalibrate the detector. |
2 | Dirty or defective flow cell. | Repeat the test with the flow cell removed. If the test is OK, exchange the flow cell components. |
3 | Dirty or defective holmium oxide filter. | Run the holmium oxide filter test. If the test fails, contact your Agilent service representative. |
4 | Optical misalignment. | Please contact your Agilent service representative. |