The Dark Current Test measures the leakage current from the sample and reference circuits. The test is used to check for defective sample or reference diodes or ADC circuits which may cause non-linearity or excessive baseline noise. During the test, the shutter is moved into the light path. Next, the leakage current from both diodes is measured.
Run the Dark Current Test with the Agilent Lab Advisor (for further information see Online-Help of user interface).