Set peakwidth as close as possible to the width (at half height) of a narrow peak of interest.
Set peakwidth as close as possible to the width (at half height) of a narrow peak of interest.
{Comment by , : ToDo: Condition: "($ProductNumber='G1314F')" }Refer to --- MISSING LINK ---.
{Comment by , : ToDo: Condition: "($ProductNumber='K1314F')" }Refer to --- MISSING LINK ---.
Choose the sample wavelength.
at a longer wavelength than the cut-off wavelength of the mobile phase,
at a wavelength where the analytes have strong absorptivity if you want to get the lowest possible detection limit,
at a wavelength with moderate absorptivity if you work with high concentrations, and
preferably where the spectrum is flat for better linearity.
Consider to use time-programming to further optimization.