A variety of flow-cell cartridges can be inserted using the same quick and simple mounting system.
The flow cells have an integrated RFID tag that contains the flow cell specific information (e.g. part number, cell volume, path length, ...). A RFID tag reader reads out this information and transfers it to the user interface.
| Standard | Semi-micro | Micro | High Pressure | Bio Standard | Bio Micro |
---|---|---|---|---|---|---|
Maximum pressure (bar (MPa)) |
40 (4) |
40 (4) |
120 (12) |
400 (40) |
|
|
Path length (mm) |
10 (conical) |
6 (conical) |
3 (conical) |
10 (conical) |
10 (conical) |
3 (conical) |
Volume (µL) |
14 |
5 |
2 |
14 |
14 |
2 |
Inlet i.d. (mm) |
0.25 |
0.17 |
0.12 |
0.25 |
14 |
0.12 |
Inlet length (mm) |
750 |
250 |
310 |
750 |
750 |
310 |
Outlet i.d. (mm) |
0.30 |
0.17 |
0.17 |
0.17 |
0.30 |
310 |
Outlet length (mm) |
120 |
120 |
120 |
120 |
120 |
310 |
Total volume (µL) |
60.77 |
14.49 |
14.00 |
60.77 |
60.77 |
310 |
Materials in contact with solvent |
SST, quartz, PTFE, PEEK |
SST, quartz, PTFE |
SST, quartz, PTFE |
SST, quartz, Kapton |
MP35N, sapphire |
MP35N, sapphire |
base-id: 3283496715