CAUTION
"Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
NOTE
The Installation of the leak tray and the external leak sensor is optional.
If you want to attach the leak tray to the valve drive you need to slide the leak tray on to the valve drive before having both clamps attached to the valve drive.
Ensure the power switch at the rear of the module is OFF and the power connector is unplugged.
Identify the marks for the screws on the side panel of the cabinet assy and screw the valve rail to the side panel of the cabinet.
Slide the first clamp on to one side of the valve drive housing.
Install the leak tray and leak sensor to the valve.
Slide the second clamp on to the same side as the first one.
Slide the valve on to the valve rail.
Attach the waste tubing to the leak plane and guide it to a proper waste container.
Connect the power cable to the power connector at the rear of the module.
Connect the CAN interface connection.
Power on the module by switching the Power button at the rear of the module if a valve head has been already installed. Otherwise continue with G7116B_Installation of the Valve Heads.