The Sensors Offset Calibration tries to find the offset between the two sensors built into the heat exchanger. The idea is to bring both sides to ambient temperature and minimize temperature difference between the heat sink and heat exchanger. This is to ensure that no temperature gradient inside the heat exchanger is present and both temperature sensors are exposed to the exact same temperature.
Because the lower sensor responds much faster to temperature changes by the Peltier, it is also important, that the heat sink is at the same temperature as the heat exchanger. Otherwise, a temperature flow from or to the heat exchanger would generate differences in temperature measured by each sensor.
Required:
After replacement of heater assembly.
After replacement of mainboard.
base-id: 3002955531