The Thermostat Test is used to evaluate the cooling and heating performance of the two Peltier elements.
The instrument starts the test when the heater temperature is within 5 °C of the current measured ambient temperature. After the start, the instrument is cooling down to 4 °C below the starting temperature. The cool-down time is used to determine the cooling performance. In the next step, the heating performance is calculated by heating up to the test starting temperature.
A typical Thermostat Test profile is shown below.