Parameter Name | External Value | Comment |
---|---|---|
Weight | 1.9 kg (4.3 lbs) |
|
Dimensions (height × width × depth) | 90 x 90 x 300 mm (3.54 x 3.54 x 11.8 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 20 VA, 4 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9843 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM classification | ISM Group 1 Class B | According to CISPR 11 |
Kit PN | Valve PN | Description | max. Pressure [bar] | Fittings | Liquid Contact | pH Range |
---|---|---|---|---|---|---|
G5641A | 5067-6682 | 2-position/10-port valve head, bio | 1300 | 10-32 | PEEK, MP35N | |
G4231A | 5067-4282 | 2-position/6-port valve head | 800 | 10-32 | PEEK, ST | |
G4231C | 5067-4241 | 2-position/6-port valve head | 1300 | 10-32 | PEEK, ST | |
G4232C | 5067-4283 | 2-position/10-port valve head | 800 | 10-32 | PEEK, ST | |
G4232D | 5067-4240 | 2-position/10-port valve head | 1300 | 10-32 | PEEK, ST | |
G4234A | 5067-4284 | 6-position/14-port valve head | 800 | M4 | PEEK, ST | |
G4234C | 5067-4273 | 6-position/14-port valve head | 1300 | M4 | PEEK, ST | |
G4235A | 5067-4159 | 12-position/13-port valve head, bio-inert | 210 | 10-32 | PEEK, Ceramic | |
G4237A | 5067-4279 | 4-position/10-port valve head | 800 | M4 | PEEK, ST | |
G4239C | 5067-4233 | 8-position/18-port valve head | 1300 | M4 | PEEK, ST | |
G5631A | 5067-4148 | 2-position/6-port valve head, bio-inert | 600 | 10-32 | PEEK, Ceramic | |
G5632A | 5067-4132 | 2-position/10-port valve head, bio-inert | 600 | 10-32 | PEEK, Ceramic | |
G5639A | 5067-4134 | 4-position/10-port valve head for columns selection, bio-inert | 600 | 10-32 | PEEK, Ceramic |
[1]Incompatible with some mineral acids.
Shipping and Storage Temperature: -40 – 80 °C
NOTE
Shipping, storing or operating this valve below 0 °C with water in the fluid passages may cause failure of the sealing surfaces.
NOTE
Operating valves above their maximum pressure limit will harm them. Please set a corresponding max. pressure limit in your Chromatographic Data System if a valve is positioned in the high pressure path.
base-id: 9163538827
id: 27021606927761803