| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 8710-2394 | Hex key 9/64 inch 15 cm long T-handle |
1 | ![]()
|
| Cleaning tissue and appropriate solvent like isopropanol or methanol |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5068-0279 | Rotor Seal, 3-position/6-port FI Valve (for G7167C) OR |
1 | ![]() | 5320-0005 | Rotor Seal, 3-position/6-port FI Valve, MP35N, 1300 bar (for G7137B) |
1 | ![]() | 5068-0280 | Stator (optional for G7167C) Replacement only necessary in case of wear OR |
1 | ![]() | 5320-0004 | Stator, Bio (optional for G7137B) Replacement only necessary in case of wear |
Prerequisites
|
CAUTION
Reduced life time of the injection valve
Component cleanliness is crucial for the life time of the injection valve.
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
In the Agilent Lab Advisor software select Service & Diagnostics > Maintenance Positions > Change Rotor Seal , and click Start.
OR
In the Local Controller start the maintenance mode and select Change Rotor Seal function.
Open the front door.
Close the front door.
In the Local Controller close Change Rotor Seal.
OR
In the Agilent Lab Advisor software Change Rotor Seal , click Back.
For G4767A, end the procedure with a Pressure Test.
For G7167C and G7137B, proceed with the subsequent steps.
Perform Hydraulic Path Leak Test.
Perform Sampler Leak Test if needed for troubleshooting in case of Hydraulic Path Leak failure.