When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 9301-0411 | Syringe, Plastic |
1 | ![]() | 9301-1337 | Syringe adapter |
1 | ![]() | G7120-44000 | Protective Foam |
Prerequisites
|
WARNING
Heavy weight
The module is heavy.
CAUTION
Mechanical damage
Flush system with appropriate storage solution, for example 20 % isopropanol in water.
Remove solvent inlets from solvent reservoirs. Disconnect the solvent tubing from the inlet of primary pump heads for both solvent channels. Use a syringe for removing liquid from the solvent tubings between solvent reservoir, shutoff valve panel, solvent selection valve, degassing unit and pump inlets.
Remove tubing and capillary connections to other modules and the solvent cabinet.
Remove the shutoff valve panel by pulling it downwards.
You may keep internal tubing and capillary connections.
Remove cable connections to other modules. Remove the module from the stack.
Carefully insert the Protective Foam to the front part of the instrument. Do not damage any tubing or capillary connections.
Close the front cover.
For transport or shipment, put the module and accessory kit in the original shipment box.