Error ID: 25463
Probable cause | Suggested actions | |
---|---|---|
1 | Pump tubing blocked. | Verify that the solvent tubing is not blocked. |
2 | Pump motor is defective. | Replace the pump motor. |
3 | Pump is defective. | Replace the peristaltic pump. |
base-id: 3803958283
id: 27021601568181259