Error ID: 25400
Cleaning procedure failed. Parameter shows what kind of cleaning procedure has failed: 1 = Wash, 2 = Prime, 3 = Autoclean, 4 = Clogged seat.
Probable cause | Suggested actions | |
---|---|---|
1 | Solvent lines not installed correctly (valve block or flushpump). | Check status of the solvent lines. Use isopropanol for verification. |
2 | Clogged needle seat. | Replace the needle seat. |
base-id: 3803778571