Error ID: 25409
The autosampler failed to complete initialization correctly. The autosampler initialization procedure can not move the motors inside of the sample handler to their reference positions in a predefined routine. During initialization, the processor monitors the position sensor and encoders to check for correct movement. If one or more of the movements or the status information of the subparts is not read out successfully, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Sample handler is blocked. | Check if the sample handler can move freely. Switch off the instrument. Do an autoreferencing (the needle assembly must be installed in the needle parkstation during this procedure). |
2 | Defective sample handler motors. | Please contact your Agilent service representative. |
base-id: 3803727371